Spectroradiometer SR-UL2
Features
-
High accuracy with spectrophotometry and high-speed measurement
-
Luminance and chromaticity in small area can be measured without attachment lens.
-
Measurement angle are selectable from 2°/ 1°/ 0.2°/ 0.1°)
-
The stable measured value are obtained even when measuring flicker light source.
-
Communication between the SR-3AR and PC is accelerated with USB and STB command.
-
Application soft CS-900A is standard accessories.
Specifications
Model
Photo detector
Dispersing element
Measurement angle
Measurement distance
Spectral band width
Wavelength accuracy
Wavelength range
Wavelength resolution
Measurement mode
Measurement object
Calculation function
Range of guaranteed illuminance accuracy *1*2
Accuracy in luminance *1
Accuracy in chromaticity *1
Repeatability in luminance *3
Repeatability in chromaticity *4
Polarization error
Measurement time *5
Interface
Power supply
Power consumption
Use conditions
Outer demensions(WxDxH)
Weight
SR-3AR
Electronically cooled linear CCD
Diffraction grating
2°/1°/0.2°/0.1° selectable (Motor moved)
350mm - ∞ (distance from metallic tip of objective lens)
5 - 8nm *Half bandwidth
±0.3nm * for Hg emission lamp
380 - 780nm
1nm
Auto/Manual(integral time/frequency) / External vertical sync signal input
Spectral radiance : W・sr-1・m-2・nm-1
Radiance Le : W・sr-1・m-2, Luminance Lv : cd・m-2,
CIE1931 chromaticity coordinates xy,
CIE1976 chromaticity coordinates u' v', Tristimulus value XYZ,
Correlated color temperature and Deviation,
CIE standard observer 2°/ 10°
2°: 0.1 - 3,000 1°: 0.3 - 9,000
0.2°: 7.5 - 70,000 0.1°: 30 - 300,000
±2% or less
±0.002
0.3% (0.1cd/m² or more)
0.0005 (0.1cd/m² or more)
Luminance 1% or less,Spectral radiance 2% ore less(400 - 700nm)
NORMAL Mode: about 1 - 31sec, HIGH SPEED Mode: about 1 - 17sec
USB2.0 / RS-232C selectable
Dedicated AC adapter AC100V-240V, 50/60Hz, DC12V
about 34W
Temperature: 5 - 35℃, Humidity: 80%RH or less (no condensation)
150×406×239mm
5.5k
(*1) : For standard illuminance A.
(*2) : Measurable range in Normal and High speed mode.
(*3) : 2σ from 10 times continuousmeasurement at measuring angle 2° in normal speed mode.
(*4) : (Max – Min) value from 10 times continuousmeasurement at measuring angle 2°in normal mode.
(*5) : Except communication time with PC.