TOPCON Digital Illuminance Meter IM-600/IM-600M

​광도(cd) 표시 가능 조도계

Mini Head type - IM-600M

Digital Illuminance Meter IM-600

조도계

Features

 

  • Wide range measurement (0.005 - 999,000 lx)

  • Display unit and light sensing unit are Separable with optional cable (IM-600)

  • Multiple measurement function

  • Response time, screen flicker can be measured with oscilloscope

  • Zero calibration function without cap can be installed equipment

Specifications

Model

JIS Class

Photo detector

Measurement range

Cosine correction

Relative spectral response

Measurement Accuracy

Measurement mode

Measurement function

Calibration function

Analog output

Interface

Power source

Operating condition

Dimensions(WxDxH)

Weight

IM-600

Conforms to the general AA class illuminance meter
(JIS C 1609-1-2006)

Silicon photo diode

0.005~999,000 lx

10° : Within ±1%
30° : Within ±1%
50°: Within ±5%
60° : Within ±5%
80° : Within ±10%

6% or less (*1)

±2% of rdg.  ±1 digit (Auto Range)

Auto/Manual, 6-range

Illuminance(lx), Illuminance difference(lx), Illuminance ratio(%)
Integrated illuminance(lx·h), Integration time(h), Intensity of light(cd)

CCF (Color Correction Factor) setting

0 - 3V

USB

Size AA dry-cell battery×2 or dedicated adopter (option)

Temperature : -10 - 40℃,
Humidity : 85% RH or less (no condensation)

188×65×34mm, Sensor Unit (IM-600M) : 16×21mm

IM-600: 200g,  IM-600M: 230g  *non-dry cell battery

Model

JIS Class

Photo detector

Measurement range

Cosine correction

Relative spectral response

Measurement Accuracy

Measurement mode

Measurement function

Calibration function

Analog output

Interface

Power source

Operating condition

Dimensions(WxDxH)

Weight

IM-600M

Conforms to the general AA class illuminance meter
(JIS C 1609-1-2006)

Silicon photo diode

0.005~999,000 lx

10° : Within ±1%
30° : Within ±2%
50°: Within ±6%
60°: Within ±7%
80°: Within ±25%

6% or less (*1)

±2% of rdg.  ±1 digit (Auto Range)

Auto/Manual, 6-range

Illuminance(lx), Illuminance difference(lx), Illuminance ratio(%)
Integrated illuminance(lx·h), Integration time(h), Intensity of light(cd)

CCF (Color Correction Factor) setting

0 - 3V

USB

Size AA dry-cell battery×2 or dedicated adopter (option)

Temperature : -10 - 40℃,
Humidity : 85% RH or less (no condensation)

188×65×34mm, Sensor Unit (IM-600M) : 16×21mm

IM-600: 200g,  IM-600M: 230g  *non-dry cell battery